Perchlorate + pH ISE Meter

ISE + Dual pH Perchlorate Meter

This laboratory precision Perchlorate ION meter is a nanotechnology instrument used to obtain reliable data using data combining a perchlorate selective electrode and a pH electrode.
Multiparameter instrument for CLO4- and pH ion concentration analysis (imacimus 2: pH and ISE)

This complete kit includes potentiometer, perchlorate ion selective electrode, pH, reference electrode and calibration standards.

A perchlorate ion-selective electrode (ISE) is used in applications where precise measurement of perchlorate ions (ClO₄⁻) is required.

Common areas of use include:

Environmental monitoring: Testing soil and water for perchlorate contamination, often near industrial sites, military areas, or agricultural lands, as perchlorate is a common contaminant.

Water treatment facilities: Monitoring perchlorate levels in drinking water to ensure compliance with health and safety standards.

Food safety and agricultural analysis: Measuring perchlorate in crops and food products, as it can accumulate in plants and affect food quality.

Research laboratories: Used in studies of perchlorate’s environmental impact, health effects, and remediation techniques.

This electrode enables rapid, selective detection of perchlorate, making it a valuable tool in both field and laboratory settings.

IC2-KIT (pH Meter + Perchlorate)

The perchlorate ion analyser, combined with the pH sensor, includes advanced analysis software that displays real-time measurements and facilitates identification. The software also allows custom calibrations with up to five set points, among other features. Continuous monitoring is available for short-term reactions or process studies. The perchlorate ion sensor is modular and can be replaced with a new perchlorate sensor or any other sensor from our catalogue. It is important to note that the ISE sensor has a limited lifetime, even when not in use, whether stored on a shelf or installed in a probe. Generally, the perchlorate sensor achieves a lifetime in excess of 12 months. The lifetime may vary between different sensors, sometimes by several months, even under the same conditions.

  • Dual Channel ION/mV/pH for use with PC/Laptop.
  • Software for Windows
  • USB cable for data/power
  • Single ION PROBE.
  • Mini Perchlorate Electrode
  • Reference electrode/pH
  • Calibration Standard 10 mg/L Perchlorate (100 mL)
  • Calibration Standard 100 mg/L Perchlorate (100 mL)
  • Calibration Standard 1000 mg/L Perchlorate (100 mL)
  • pH buffers 4/7 (2x50ml)
  • Deionized water (50 mL)
  • User manual

Your Perchlorate Ion and pH tool kit comes with all the items you need to start reading samples.

Employing the same operating principle as standard laboratory electrodes.

  • The potentiometry technique (ISE) is relatively inexpensive and easy to operate.
  • Ion selective electrodes have a wide concentration measurement range.

Specifications Data
ISE range 0.4…8.000 mg/l
ISE/pH resolución 0 to 14pH es 0,01 pH
0 to 1.0 ppm es 0.1 ppm
1 to 9000 ppm es 1 ppm
ISE accuracy ±4% * In the absence of interfering ions, the accuracy can be better than ±1mV, which corresponds to ±4% of the measurement.
ISE repeatability ±10% monovalent ** Repeatability in the measurement of a standard solution after calibration
ISE /pH calibration 2 to 5 points. Fully customizable calibration standards.
Power Supply USB – 5V -0.5A
Temperature range 4- 50 ºC
Interface Imacimus software application for Windows
Connectivity Mini USB female AB type connector
Method Electrochemical / ISE

imacimus 2 guide

Guidance software

Single ION guide

Click here to visit the warranty page

Additional information

Weight 2 kg